My DUT Has a Different Sampling Rate Than My Audio Interface. Is There a Way That I Can Analyze Recordings Made on My DUT in SoundCheck?

Analysis in SoundCheck requires that the stimulus waveform and the recorded time waveform have the same sampling rate. SoundCheck contains two post-processing operations that can convert the waveform from your DUT to the proper sampling rate for analysis. These post-processing functions are Resampling and Frequency Shift which should be used any time a waveform is recorded or played back through an external device and is brought back into SoundCheck for analysis.

The Bluetooth Headset example sequence that comes with SoundCheck shows how to use these steps in a sequence. 